Installation
*INSTALLATION*
147
Securing to the
worktop
To improve stability the machine must be secured to the worktop
after it has been aligned.
Open the door.
Screw the machine to the continuous worktop through the holes in
the front trim on the left and right.
Please contact the Miele Customer Service Department to secure it
at the sides to adjacent cabinetry.
Venting the
circulation pump
The gaps between a built-in machine and adjacent cabinetry
must not be sealed, e.g. with silicone sealant, as this could
compromise ventilation of the circulation pump.
Protective foil/
Worktop
protector
The protective foil supplied protects the worktop from damage
caused by steam when the door is opened. It should be positioned
underneath the worktop above the machine door.
Electromagnetic compatibility (EMC)
The machine has been tested for electromagnetic compatibility (EMC)
in accordance with EN61326-1 and is suitable for operation in
commercial environments, such as hospitals, medical practices and
laboratories and other similar environments which are connected to
the mains power supply.
The machine‘s high frequency (HF) energy emissions are very low and
are therefore unlikely to interfere with other electronic machines in the
vicinity.
Flooring in the installation area must be wood, concrete or tiled.
Synthetic flooring must be able withstand a relative humidity level of
30% to minimise the risk of electrostatic discharges.
The quality of the power supply should comply with that found in a
typical commercial or hospital environment. Check that the power
supply voltage is within a range of +/-10% of its nominal value.