Dispensing calibration
Dispensing calibration is used to measure the throughput of the indi-
vidual dispenser pumps.
To go directly to
Dispensing cali-
bration when the
machine is
switched off:
Press and hold multifunction button2 and switch the machine on
with “Power on”.
Dispensing calibration Main menu
Dispensing pump 1
Dispensing pump 2
Options:
- Dispenser pump 1to13
Select a dispenser pump.
When a dispenser pump is selected, the following will appear in the
display:
100 ml
Dispensing calibration Dispensing pump 1
V:
P:
t:
s
No
0
0/1
Tip: The dispensing quantity can be set from 0to9,999ml in 1ml in-
crements.
Press the multifunction button0/1.
The dispensing hose will fill up.
Stop the process by pressing the 0/1button.
Run the hose of the dispenser pump upwards into a measuring ves-
sel.
Hold the entire measuring vessel above the dispenser pump while
doing this.
Irritant detergents.
Skin and eye irritation.
Avoid contact with skin and eyes, wear protective clothing.
Return to the main menu by pressing the “Back”button.
Dispensing calibration Main menu
Dispensing pump 1
Dispensing pump 2
Select a dispenser pump.