Productionconguration
The system includes a periodic EDI rinsing feature to ensure a good system performance and
water quality is always maintained.
The rinsing will take place at the time of the day set in Periodic EDI rinsing check time, that
can be chosen by the system manager to better adapt to the work time schedule.
The system will rinse the EDI for the Periodic EDI rinsing duration (1h by default) whenever
the system did not produce Elix water for at least 1 hour during the time period set in the
Periodic EDI rinsing period (24h by default).
PODconguration
This app allows a user to congure all parameters specic to PODs. These are unique to the
POD that is being used to input the values.
To duplicate POD parameters, the action(s) should be repeated on all other PODs.
POD Name
This can be personalized. Click on the text box and input up to a maximum of 8 characters.
Screen brightness
Adjust to the preferred brightness from 1 to 7 using the arrows or click on the box to access
the keypad and type-in the value.
Sound volume
Each POD can emit a sound when alerts/alarms are triggered. This can be activated or deactivated
using the slider button. The sound can be adjusted to the desired volume by using the arrows or
click on the box to access the keypad and type-in the value. By default, the sound is deactivated.
Flowmeter oset
The owmeter has been calibrated in order to achieve volume precision of +/- 2% for Q-POD and
+/- 5% for E-POD. Laboratory glassware accuracy can vary a lot. This setting provides a user
with the ability to adapt the owmeter to their current glassware accuracy by using an oset
function.
Adjust the oset by using the arrows or click on the box to access the keypad and type in the
value. To gain back the original calibration settings, set the value back to 0.