Settings
23
Alarmconguration
Adjust the alarm set points by either clicking on the arrows or alternatively, click in the box to
access the keypad and directly type-in the value.
Water quality
• Product resistivity/conductivity
• Product TOC
Inlet strainer clean
It is recommended to clean the inlet strainer once a year to avoid it becoming clogged. Depend-
ing on the feed water and its particle concentration, the inlet strainer clean frequency can be
adapted.
Cartridge lifetimes
Cartridges should be changed when the system alerts the user.
Saturated pack content will start to let ion and organic contaminants through. This is called
breakthrough and often starts with traces that cannot be detected by on-board monitoring.
Some technologies will clog, creating a back pressure which can result in owrate issues or pres-
sure alarms in the system.
In validated settings, to adjust lifetimes according to existing SOPs, click on the applicable
lters and adjust accordingly.