Self Help Guide
33
SELF HELP GUIDE
Depressurize system
Depressurize both the Production and Distribution sections of the system in case of water
leak to temporarily stop production and all water processes. Go to page 32 Maintenance >
Depressurization.
Identify the source of the leak. Once xed, to exit this mode, a user will need to go back to
the home screen on which the initial depressurization took place and select Exit button on the
drop-down screen for dispensing to become available again on all Q-PODs.
Turn system o
Press the ON/OFF switch located on
the left side of the unit.
Note: When the system is on, the
power switch is lit.
& UNPLUG
Important! Never open the system, even when the power switch has been switched o.
Hazardous voltage is present with a hazard of electrical arc ash. Will cause death or serious
injury.
It is not sucient to stop all power to system, the plug must be physically removed from its power
source. Installation and maintenance should only be done by a qualied person. Appropriate
personal protective equipment (PPE) must be worn and safe work practices must be followed.
If the system is powered o for 20 days, the capacitor will completely discharge. When
powering on again the system, the time & date settings will need to be reset, see page 24
Information>Date Time & Language. It takes about 45 minutes to recharge the capacitor.
Inaccurate volumetric dispensing
Volumetric dispensing has been validated to work with a +/- 2% precision for Q-POD and +/-
5% precision for E-POD. Lab glassware precision can vary and to accommodate local needs a
ow meter oset menu has been created. Go to page 21 Settings>Flow meter oset for more
information.
High TOC values
TOC levels can vary depending on the type of feed water or the laboratory environment.
Here are a few recommendations to follow:
1. A pack replacement can temporarily increase the TOC until they have been completely
rinsed out.
2. Clean the A10 TOC Monitor cell, go to page 31 Maintenance>A10 Cleaning