Self Help Guide
34
Low owrate
Ensure the POD-Pak is not air-locked. This can happen in the weeks that follow a
replacement of IPAK cartridges. Dispense water and open the POD-Pak air vent to see if
there is any trapped air. Close the vent once the air has been released.
One possible reason for a decrease in Milli-Q Water owrate is a clogged POD-Pak. The
POD-Pak should be replaced when it appears to be clogged.
Tank is not re-lled of pure water
Open the drop down menu swiping down on the touchscreen to check the status of the
pure production section.
Exit the maintenance or rinsing mode in case it is activated.
Solve any alarms that could be stoping the production process.
Contact US
Get in contact with the local technical support hotline. The details can be found in the contacts
app, go to page 14 Information >Contacts or visit our website www.emdmillipore.com (North
America) or www.merckmillipore.com (Rest of the World).