5-4
4. As shown in the figure: Remove the cover of the push and pull box and then remove
the drive gear (Note: The pressure sensor and the board cannot be removed).
5. As shown in the figure: Remove the pressure sensor wire and the hollow shaft outlet
from the pressure detection board PCBA, and then pull the push and pull box out of
the equipment.
Gear