5-5
5.2.3 Removing the Pump Body Assembly (Note: You should
wear an anti-static wristband throughout)
1. As shown in the figure: Remove the four ST3x6 Phillips head self-tapping screws
with a Phillips screwdriver.
2. As shown in the figure: Use a Phillips screwdriver to remove the ST3x10 Phillips
head self-tapping screws that fix the pump body.
Cut the cable tie.