5-3 
6.  Verify the realtime CO
2 
value is within 6.0 ± 0.3% in the [Calibrate CO2] menu. 
 
5.2.2.3 Calibration 
Tools required: 
  A steel gas cylinder with 5±0.03% CO
2
, 21.0% O
2
 and balance gas N
2
 (P/N 0075-00-0033-01), or a steel gas cylinder 
with: 
  CO
2
 concentration 3% - 7% 
  a/c  ≤  0.01 (where a = absolute gas concentration accuracy, c = gas concentration) 
  balance gas N
2
 
  T-shape connector 
  Tubing 
 
To perform a calibration: 
1.  Make sure that the sidestream or microstream CO
2
 module is warmed up. 
2.  Check the airway for leaks. 
3.  Select [Main Menu] 
→
 [Maintenance >>] 
→
 [User Maintenance >>] 
→
 enter the required password 
→
[Module 
Maintenance>>]→[Maintain CO2 >>] 
→
 [Calibrate CO2 >>]. 
4.   In the [Calibrate CO2] menu, select [Zero]. 
5.  After the zero calibration is finished successfully, connect the equipment as follows:   
 
6.  Open the valve to flow CO2 and make sure that there is flow sufficient to vent to atmosphere. 
7.   In the [Calibrate CO2] menu, enter the CO2 concentration in the [CO2] field. 
8.   In the [Calibrate CO2], the measured CO2 concentration is displayed. After the measured CO2 concentration 
becomes stable, select [Calibrate CO2] to calibrate the CO2 module.   
 
If the calibration is finished successfully, the message [Calibration Completed!] is displayed in the [Calibrate CO2] 
menu. If the calibration failed, the message [Calibration Failed] is displayed. In this case, perform another calibration. 
Monitor 
Cylinder 
Flowmeter 
Relief valve 
T-shape connector 
Sampling line