5-24 Image Optimization 
Speed 
Description  This function is used to set the scanning speed of PW mode imaging. 
Operations  Rotate the knob under the [Speed] item on the touch screen. 
There are 6 levels of scan speed available: the smaller the value the faster 
the speed. 
Effects  Changing the speed makes it easier to identify the cardiac cycles and to 
detect more details. 
 
T/F Res. 
Description  This function is used to create a balance between time resolution and 
spatial resolution. 
Operations  Adjust using the [T/F Res.] item on the touch screen. 
There are 0-4 levels of T/F Res. values available. 
 
Wall Filter 
Description  It filters out low-velocity signals to provide effective information, and this 
function is used to adjust the filtered frequency. The real-time value is 
displayed in the image parameter area in the top-right corner of the 
screen. 
Operations  Select using the [WF] item on the touch screen. 
0-6 levels of wall filter function are provided. 
Impacts  Signals of low-velocity flow may be missing. 
 
Tint Map 
Description  This function provides an imaging process based on color difference 
rather than gray distinction. 
Operations  Adjust through the [Tint Map] item on the touch screen to select the map. 
There are 1-25 color effect maps available. Off represents the function is 
disabled. 
 
Gray Map 
Description  This function applies the gray correction to obtain optimum images. 
Operations  To select from among the maps, turn the knob under [Gray Map] on the 
touch screen. There are 1-25 gray effect maps available.