What to do if Mindray DC-70 IO interface system has LCD works normally, but NO video printer output when pressing ?
C
Cheryl BlackburnSep 15, 2025
If the LCD works normally on your Mindray Medical Equipment, but there is no video printer output when pressing on the control panel, it could be due to a USB connection error or a PC carrier board error. Re-plug the USB. Plug it into another USB port to confirm. Use a U disk to confirm whether other USBs work well. If the USB port is abnormal, replace the PC carrier board.
W
William FisherAug 25, 2025
What to do if the Mindray Medical Equipment AC indicator is on, the power on indicator blinks green, and the power status indicator (3.3 V/5 V/12 V) is off?
J
John CobbAug 25, 2025
If the AC indicator remains on, the green power on indicator blinks after pressing the power button, and the power status indicator (3.3 V/5 V/12 V) is off, the PC module is likely the cause. Replace the PC module.
D
Daniel RobertsAug 25, 2025
What to do if the Mindray DC-70 AC indicator is on and the power on indicator blinks yellow?
W
william46Aug 26, 2025
If the AC indicator remains on and the power on indicator starts to blink yellow after pressing the power button, the DC-DC board likely fails to respond. Replace the DC-DC board.
C
Cindy MejiaAug 26, 2025
How to troubleshoot a Mindray Medical Equipment when the AC power indicator remains off?
L
Lindsey CordovaAug 27, 2025
If the AC power indicator remains off, it could be due to no AC input or errors on the AC-DC assembly. Check the AC connection. If the issue persists, replace the AC-DC assembly.
A
Amanda KimSep 16, 2025
What to do if Mindray DC-70 control panel has some or a certain key on the small keyboard fails?
J
James ReyesSep 16, 2025
If some or a certain key on the small keyboard fails on your Mindray Medical Equipment control panel, replace the small keyboard assembly.
V
vgutierrezSep 23, 2025
What to do if Mindray Medical Equipment system cannot power on?
J
Joshua SandersSep 23, 2025
If the Mindray Medical Equipment system fails to power on, begin by checking the backlight of the power button. Also, inspect the AC indicator on the control panel and verify the three power supply indicators on the IO back board.
M
Maria JonesSep 7, 2025
What to do if Mindray DC-70 Medical Equipment IO interface system fails to recognize Wi-Fi?
S
Steven WilsonSep 8, 2025
If your Mindray Medical Equipment fails to recognize Wi-Fi, check the 3.3 V indicator on the IO board. If it's brighter or off, the PC carrier board may be the issue, so replace it. If the network and settings are correct, the Wi-Fi board or PC carrier board may be faulty, so replace the Wi-Fi board or PC carrier board.
S
Sarah CalhounSep 8, 2025
What to do if Mindray DC-70 has no echo in ultrasound image?
R
rebecca72Sep 9, 2025
If there is no echo in the ultrasound image on your Mindray Medical Equipment, even though the probe is recognized, it could be due to a PHV voltage output of the power module being 0 V or abnormal, or errors on the TR64 board, DC-DC board, or probe board. Confirm the error by replacing the TR64 board, DC-DC board, or probe board.
A
Amber HughesSep 2, 2025
What causes dark strips on B image in Mindray Medical Equipment?
W
wgardnerSep 2, 2025
Dark strips appearing on the B image of your Mindray Medical Equipment can be caused by a probe malfunction, such as array corruption. Replace the probe. Alternatively, replace the TR64 board. If the strip is not fixed well for a single probe, the probe board may be faulty, so replace the probe assembly.
J
Jeffrey LarsonAug 31, 2025
What to do if Mindray DC-70 Medical Equipment system cannot start and BIOS start-up graphics is normally displayed, but it does not last?
S
Stephanie WoodsAug 31, 2025
If the BIOS start-up graphics are normally displayed on your Mindray Medical Equipment, but it does not last, it indicates a PC module failure. Replace the PC module.
Mindray DC-70 Specifications
General
Power Supply
100-240V AC, 50/60Hz
Imaging Modes
B-mode, M-mode, Color Doppler, Power Doppler, Pulsed Wave Doppler, Continuous Wave Doppler
Probes
Convex, Linear, Phased Array, Volume
Connectivity
USB, Ethernet, VGA
Advanced Features
3D/4D Imaging
Applications
Abdominal, Obstetrics, Gynecology, Cardiac, Small Parts, Vascular, Musculoskeletal, Urology