Page #210 background image
11-10 System Diagnosis and Support
11.3.4 Fan Error
Alarming tips LOG record Suggestion
Please
contact
service
engineers if
fans need
maintenance.
*** System Monitor: Fan alert[XXX], Current speed : [VVV] rpm, Limit
speed: [LLL] rpm
[XXX] represents fan name; [VVV] represents the current value; and
[LLL] represents the limit.
The names for the fans are:
FS_TOP_TACH0, FS_TOP_ TACH1, FS_TOP_ TACH2, FS_DOWN_
TACH0, FS_DOWN_ TACH1, FS_DOWN_ TACH2, FS_PCThe positions
of the fans in the device:
Replace the
fan/re-connect
the
circuit/clear
off fan
blocking.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray DC-70 and is the answer not in the manual?

Mindray DC-70 Specifications

General IconGeneral
Power Supply100-240V AC, 50/60Hz
Imaging ModesB-mode, M-mode, Color Doppler, Power Doppler, Pulsed Wave Doppler, Continuous Wave Doppler
ProbesConvex, Linear, Phased Array, Volume
ConnectivityUSB, Ethernet, VGA
Advanced Features3D/4D Imaging
ApplicationsAbdominal, Obstetrics, Gynecology, Cardiac, Small Parts, Vascular, Musculoskeletal, Urology
WeightApprox. 85 kg
Operating SystemWindows-based