What to do if Mindray Medical Equipment control panel keys work well, but single TGC button fails?
R
Rick MoraSep 4, 2025
If the control panel keys on your Mindray Medical Equipment work, but a single TGC button fails, it indicates a TGC board error. Replace the TGC board.
M
Michele McculloughSep 3, 2025
What to do if Mindray Medical Equipment system cannot start and System Doppler start-up graphics appears but it does not last?
B
Brett BraunSep 3, 2025
If the System Doppler start-up graphics appear on your Mindray Medical Equipment but do not last, it indicates a failure to load the system software. The hard disk or system software may be corrupted. Try to restore the operating system. If this fails, replace the hard disk.
A
Amber HughesSep 2, 2025
What causes dark strips on B image in Mindray Medical Equipment?
W
wgardnerSep 2, 2025
Dark strips appearing on the B image of your Mindray Medical Equipment can be caused by a probe malfunction, such as array corruption. Replace the probe. Alternatively, replace the TR64 board. If the strip is not fixed well for a single probe, the probe board may be faulty, so replace the probe assembly.
L
leonelizabethSep 1, 2025
How to troubleshoot Mindray DC-70 Medical Equipment if it cannot power on, AC indicator is on, power on indicator blinks green, and power status indicator is off?
P
Patrick McknightSep 1, 2025
If your Mindray Medical Equipment cannot power on with the AC indicator on, the power on indicator blinking green, and the power status indicator off, the PC module is likely failing. Replace the PC module.
J
Jeffrey LarsonAug 31, 2025
What to do if Mindray DC-70 Medical Equipment system cannot start and BIOS start-up graphics is normally displayed, but it does not last?
S
Stephanie WoodsAug 31, 2025
If the BIOS start-up graphics are normally displayed on your Mindray Medical Equipment, but it does not last, it indicates a PC module failure. Replace the PC module.
M
Marissa MartinezAug 30, 2025
Why Mindray DC-70 probe cannot be recognized?
R
Robin IbarraAug 30, 2025
If a probe cannot be recognized by the Mindray Medical Equipment, and other probes work well in all sockets, the probe itself may be malfunctioning, so replace the probe. If the probe isn't recognized when connected to a specific socket, the probe board may be malfunctioning, so replace the probe assembly. If no probe is recognized on any socket, the system can't read information from the probe board ID, indicating a probe board malfunction, so replace the probe assembly.
P
perezlatashaAug 29, 2025
What to do if Mindray DC-70 system cannot start but BIOS start-up graphics appears, and then becomes blank?
M
Mark SparksAug 29, 2025
If the BIOS start-up graphics appear on your Mindray Medical Equipment but then the screen goes blank, it means the system is unable to load, possibly because the system directory of the hard disk is corrupted. Try to restore the operating system. If that doesn't work, replace the hard disk.
A
Amy MckayAug 28, 2025
What to do if Mindray DC-70 image has noise?
R
Ruben MooreAug 28, 2025
If noise appears in the B image on your Mindray Medical Equipment, it could be due to a probe malfunction. Try replacing the probe. Another cause could be interference from other electrical equipment. Turn off electrical equipment connected to the system to check if the ripple-shaped interference signal disappears from the image.
M
Melanie HudsonAug 27, 2025
How to fix Mindray DC-70 when it cannot power on and the AC power indicator remains off?
H
Heather PalmerAug 28, 2025
If your Mindray Medical Equipment system cannot power on and the AC power indicator remains off, it could be due to no AC input or errors on the AC-DC assembly. Check the AC connection. If the problem persists, replace the AC-DC assembly.
C
Christopher SmithAug 27, 2025
What to do if Mindray Medical Equipment system cannot start and “OPERATING SYSTEM NOT FOUND” appears?
C
cgoodAug 27, 2025
If the Mindray Medical Equipment displays “OPERATING SYSTEM NOT FOUND”, it indicates that the loading operation has failed because the hard disk or the root directory of the hard disk in the operating system is corrupted. You should try to restore the operating system. If this fails, replace the hard disk.
Mindray DC-70 Specifications
General
Power Supply
100-240V AC, 50/60Hz
Imaging Modes
B-mode, M-mode, Color Doppler, Power Doppler, Pulsed Wave Doppler, Continuous Wave Doppler
Probes
Convex, Linear, Phased Array, Volume
Connectivity
USB, Ethernet, VGA
Advanced Features
3D/4D Imaging
Applications
Abdominal, Obstetrics, Gynecology, Cardiac, Small Parts, Vascular, Musculoskeletal, Urology