What to do if Mindray DC-70 control panel has some or a certain key on the small keyboard fails?
J
James ReyesSep 16, 2025
If some or a certain key on the small keyboard fails on your Mindray Medical Equipment control panel, replace the small keyboard assembly.
M
Mary ConnerSep 16, 2025
What to do if Mindray DC-70 Medical Equipment IO interface system has two loudspeakers are mute?
T
timothythomasSep 17, 2025
If two loudspeakers are mute on your Mindray Medical Equipment and the self-test for Audio Test fails, it indicates a PC carrier board error. Replace the PC carrier board.
C
Carl FisherSep 18, 2025
What to do if Mindray Medical Equipment IO interface system has MIC malfunction?
D
David ManningSep 18, 2025
If the MIC malfunctions on your Mindray Medical Equipment and the self-test for the Microphone Interface Test fails, and the MIC works well, the error may occur to the PC carrier board. Replace the PC carrier board.
T
tcervantesSep 18, 2025
What to do if Mindray DC-70 Medical Equipment ECG module has abnormal waveform?
J
Jerry BlackSep 18, 2025
If you enable ECG on your Mindray Medical Equipment and the waveform appears as a signal line or the shape of the wave is abnormal, it may be due to an abnormal ECG lead or ECG module. Self test for ECG module and replace the ECG lead or ECG module.
A
adrianelliottSep 23, 2025
What to do if Mindray DC-70 Medical Equipment IO interface system has only one loudspeaker is mute?
B
Brenda CastilloSep 23, 2025
If only one loudspeaker is mute on your Mindray Medical Equipment and the self-test for Audio Test fails, it indicates a loudspeaker error. Replace the loudspeaker.
C
Carol RojasSep 23, 2025
What to do if Mindray DC-70 Medical Equipment system cannot power on, AC indicator is on, power on/off indicator is off, and power status indicator is off?
A
Andrea DanielsSep 23, 2025
If your Mindray Medical Equipment system cannot power on, the AC indicator remains on, the power on/off indicator is off, and the power status indicator is off, the PC carrier board is likely failing. Replace the PC board.
V
vgutierrezSep 23, 2025
What to do if Mindray Medical Equipment system cannot power on?
J
Joshua SandersSep 23, 2025
If the Mindray Medical Equipment system fails to power on, begin by checking the backlight of the power button. Also, inspect the AC indicator on the control panel and verify the three power supply indicators on the IO back board.
D
David FarrellAug 23, 2025
What to do if the Mindray DC-70 AC indicator remains on, the power on indicator is off or doesn't blink, and the power status indicator (3.3 V/5 V/12 V) is off?
N
Neil BoydAug 23, 2025
If the AC indicator remains on, the power on/off indicator is off or does not blink after pressing the button, and the power status indicator (3.3 V/5 V/12 V) is off, the PC carrier board is likely the cause. Replace the PC board.
M
Martin GreenSep 6, 2025
What to do if Mindray Medical Equipment IO interface system has network disconnected or abnormal?
R
Renee WhiteSep 6, 2025
If the network is disconnected or abnormal on your Mindray Medical Equipment, and the network and settings are correct, the issue may be with the IO board or PC module. Replace the IO board or PC module.
L
Linda LewisSep 5, 2025
What to do if Mindray Medical Equipment control panel keys work well, but single encoder does not work?
G
grayalexanderSep 5, 2025
If the control panel keys on your Mindray Medical Equipment work well, but a single encoder does not work, it indicates a single encoder error. Replace the encoder board.
Mindray DC-70 Specifications
General
Power Supply
100-240V AC, 50/60Hz
Imaging Modes
B-mode, M-mode, Color Doppler, Power Doppler, Pulsed Wave Doppler, Continuous Wave Doppler
Probes
Convex, Linear, Phased Array, Volume
Connectivity
USB, Ethernet, VGA
Advanced Features
3D/4D Imaging
Applications
Abdominal, Obstetrics, Gynecology, Cardiac, Small Parts, Vascular, Musculoskeletal, Urology