How to troubleshoot a Mindray Medical Equipment when the AC power indicator remains off?
L
Lindsey CordovaAug 27, 2025
If the AC power indicator remains off, it could be due to no AC input or errors on the AC-DC assembly. Check the AC connection. If the issue persists, replace the AC-DC assembly.
D
Daniel RobertsAug 25, 2025
What to do if the Mindray DC-70 AC indicator is on and the power on indicator blinks yellow?
W
william46Aug 26, 2025
If the AC indicator remains on and the power on indicator starts to blink yellow after pressing the power button, the DC-DC board likely fails to respond. Replace the DC-DC board.
W
William FisherAug 25, 2025
What to do if the Mindray Medical Equipment AC indicator is on, the power on indicator blinks green, and the power status indicator (3.3 V/5 V/12 V) is off?
J
John CobbAug 25, 2025
If the AC indicator remains on, the green power on indicator blinks after pressing the power button, and the power status indicator (3.3 V/5 V/12 V) is off, the PC module is likely the cause. Replace the PC module.