C-10 Description of Self Test Items
Suggestion to failure test
Restart the device to perform the self test if the test result appears Error. It is necessary to restore
the device (OS+doppler) if Error re-appears.
It is recommended to replace TRA board if the test result is FAIL.
C.1.18 DSP FPGA and TRA Interconnection Test (Data
Interface)
Top test item
DSP FPGA and TRA FPGA communication self test (control interface)
Test content
Test whether XCVER data interface between TRA FPGA on the main board and DSP FPGA works
well.
Analysis to test failure
The drive goes wrong if the test result appears Error.
XCVER data interface on the connection of TRA FPGA-DSP FPGA goes wrong if the test result is
FAIL.
Suggestion to failure test
Restart the device to perform the self test if the test result appears Error. It is necessary to restore
the device (OS+doppler) if Error re-appears.
It is recommended to replace TRA board if the test result is FAIL.
C.1.19 PC Module and DSP FPGA Interconnection Test
Top test item
N/A
Test content
Test whether PCIe interface communication between PC module and DSP FPGA works well.
Analysis to test failure
Check whether Windows device manager recognizes Ultrasound Backbone Device. If the drive is
not installed properly, it cannot be recognized.
The drive goes wrong if the test result appears Error.
PC module and DSP FPGA has communication error if the test result is FAIL.
Suggestion to failure test
It is recommended to restore the device (OS+doppler) if the drive is not recognized.
Restart the device to perform the self test if the test result appears Error. It is necessary to restore
the device (OS+doppler) if Error re-appears.
It is recommended to replace engine board if the test result is FAIL.
C.1.20 DSP FPGA DDRIII Test
Top test item
PC module and DSP FPGA interconnection test

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray DC-70 and is the answer not in the manual?

Mindray DC-70 Specifications

General IconGeneral
Power Supply100-240V AC, 50/60Hz
Imaging ModesB-mode, M-mode, Color Doppler, Power Doppler, Pulsed Wave Doppler, Continuous Wave Doppler
ProbesConvex, Linear, Phased Array, Volume
ConnectivityUSB, Ethernet, VGA
Advanced Features3D/4D Imaging
ApplicationsAbdominal, Obstetrics, Gynecology, Cardiac, Small Parts, Vascular, Musculoskeletal, Urology
WeightApprox. 85 kg
Operating SystemWindows-based

Related product manuals