Description of Self Test Items C-17
Analysis to test failure
If the test result is FAIL, it indicates the checksum of the primary LCD monitor is not consistent
with that in the configuration file.
Suggestion to failure test
Restart the ultrasound system if the test result is FAIL.
C.1.33 Primary LCD Monitor Brightness Control Test
Top test item
I2C communication test of primary LCD monitor
Test content
The dialog box appears when performing the test. The program controls the brightness of primary
LCD monitor via 12C interface. The brightness goes down to the lowest, and then goes up to the
highest, then gets back to normal. If backlight changes, click “Yes, I Can”, otherwise click “No, I
Can’t”. If the user clicks “Retry” the system will repeat the procedure mentioned above.
Analysis to test failure
The drive goes wrong if the test result appears Error.
Brightness control of LCD monitor goes wrong if the test result is FAIL.
Suggestion to failure test
Restart the device to perform the self test if the test result appears Error. It is necessary to restore
the device (OS+doppler) if Error re-appears.
It is recommended to replace primary LCD monitor if the test result is FAIL.
C.1.34 Primary LCD Monitor Backlight Control Test
Top test item
I2C communication test of primary LCD monitor
Test content
The dialog box appears when performing the test. The program turns the backlight of primary LCD
monitor off for 3 seconds, and then goes back to normal. If backlight changes, click “Yes, I Can”,
otherwise click “No, I Can’t”. If the user clicks “Retry” the system will repeat the procedure
mentioned above.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray DC-70 and is the answer not in the manual?

Mindray DC-70 Specifications

General IconGeneral
Power Supply100-240V AC, 50/60Hz
Imaging ModesB-mode, M-mode, Color Doppler, Power Doppler, Pulsed Wave Doppler, Continuous Wave Doppler
ProbesConvex, Linear, Phased Array, Volume
ConnectivityUSB, Ethernet, VGA
Advanced Features3D/4D Imaging
ApplicationsAbdominal, Obstetrics, Gynecology, Cardiac, Small Parts, Vascular, Musculoskeletal, Urology
WeightApprox. 85 kg
Operating SystemWindows-based

Related product manuals