Image Optimization 5-5 
TGC 
Description  The system compensates the signals from deeper tissue by segments to 
optimize the image. 
There are 8-segment TGC sliders on the control panel corresponding to 
the areas of the image. 
Operations  To increase the gain compensation in an area of interest, move the TGC 
slider to the right. 
To decrease the gain compensation in the corresponding area of interest, 
move the TGC slider to the left. 
About 1.5 seconds after the adjustment is complete, the TGC curve 
disappears. 
Effects  Adjust the signal gain for a particular image area to achieve a balanced 
image. 
 
Acoustic Power 
Description  Refers to the power of ultrasonic waves transmitted by the probe, the real-
time value of which is displayed in the upper part of the screen. 
Operations  Adjust the acoustic power through the [A.Power] item on the touch screen. 
The adjusting range is 3.2-100%. 
Effects  Generally, increasing the acoustic power will increase the brightness and 
contrast of the image and the force of penetration. 
Impacts  You should perform exams according to actual situations and follow the 
ALARA Principle. 
 
Focus 
Description 
Refers to adjusting the focus of the ultrasonic beams, using the ā
ā 
symbols which are displayed to the right of the image. 
Operations  Adjust the focus number using the [Focus Number] item on the touch 
screen. 
Use the <Focus> deflector rod in the bottom-right part of the control panel 
to adjust the focus position. 
In B mode, the available focus number can be switched between 1-4. 
Effects  The area that is focused will be of a higher contrast and resolution. 
Impacts  The greater the number of focal zones, the slower the frame rate.