Guidance and Manufacturer's Declaration
D-2
TABLE 1
GUIDANCE AND MINDRAY DECLARATION—ELECTROMAGNETIC EMISSIONS
The system is intended for use in the electromagnetic environment specified below. The
customer or the user of system should assure that it is used in such an environment.
EMISSIONS TEST COMPLIANCE
ELECTROMAGNETIC ENVIROMENT
GUIDANCE
RF emissions
CISPR 11
Group1
The system uses RF energy only for its internal
function. Therefore, its RF emissions are very
low and are not likely to cause any interference in
nearby electronic equipment.
RF emissions
CISPR 11
Class B
system is suitable for use in all establishments
including domestic establishments and those
directly connected to the public low-voltage
power supply network that supplies buildings
used for domestic purposes
Harmonic
Emissions
IEC61000-3-2
Class A
Voltage
Fluctuations/Flicker
Emissions
IEC61000-3-3
Compliance