Guidance and Manufacturer's Declaration 
 
D-4  
TABLE 3 
GUIDANCE AND MINDRAY DECLARATION—ELECTROMAGNETIC IMMUNITY 
The system is intended for use in the electromagnetic environment specified below.   The 
customer or the user of system should assure that it is used in such an environment. 
IMMUNITY 
TEST 
IEC 60601-1-2 
TEST LEVEL 
COMPLIANCEL
EVEL 
ELECTROMAGNETIC 
ENVIRONMENT-GUIDANCE 
Conduced 
RF IEC 
1000-4-6 
3 Vrms   
150kHz to 
80MHz 
1 Vrms 
Portable and mobile RF 
communications equipment should be 
used no closer to any part of system, 
including cables, than the 
recommended separation distance 
calculated from the equation applicable 
to the frequency of the transmitter. 
Recommended separation distance 
d = 3.5 x
P
 
d = 1.2 x
P
  80 MHz to 800 MHz 
d = 2.3 x
P
  800 MHz to 2.5GHz 
where P is the maximum output power 
rating of the transmitter in watts (W) 
according to the transmitter 
manufacturer and d is the 
recommended separation distance in 
meters (m).b   
Field strengths from fixed RF 
transmitters, as determined by an 
electromagnetic site survey, a should be 
less than the compliance level in each 
frequency range b   
Interference may occur in the vicinity of 
equipment marked with the following 
symbol: 
 
Radiated 
RF IEC 
1000-4-3 
3 V/m   
80MHz to 2.5 
GHz 
3V/m 
Note 1    At 80 MHz and 800 MHz, the higher frequency range applies. 
Note 2  These guidelines may not apply in all situations. Electromagnetic propagation is 
affected by absorption and reflection from structures, objects and people. 
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) 
telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV 
broadcast cannot be predicted theoretically with accuracy.  To assess the 
electromagnetic environment due to fixed RF transmitters, an electromagnetic site 
survey should be considered.  If the measured field strength in the location in which 
system is used exceeds the applicable RF compliance level above, system should be 
observed to verify normal operation.    If abnormal performance is observed, additional 
measures may be necessary, such as reorienting or relocating system.