EasyManuals Logo

Mindray Eagus R9 User Manual

Mindray Eagus R9
365 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #324 background imageLoading...
Page #324 background image
A-2
Electrical Safety Inspection
ELECTRICAL SAFETY INSPECTION
1- Power Cord Plug
TEST PROCEDURE
The Power Plug
The Power Plug Pins No broken or bent pin. No discolored pins.
The Plug Body No physical damage to the plug body.
The Strain Relief
No physical damage to the strain relief. No plug
warmth for device in use.
The Power Plug No loose connections.
The Power Cord
The Power Cord
No physical damage to the cord. No
deterioration to the cord.
--For devices with detachable power cords,
inspect the connection at the device.
--For devices with non-detachable power cords,
inspect the strain relief at the device.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray Eagus R9 and is the answer not in the manual?

Mindray Eagus R9 Specifications

General IconGeneral
BrandMindray
ModelEagus R9
CategoryDiagnostic Equipment
LanguageEnglish

Related product manuals