C-4 Description of Self Test Items
Analysis to test failure
If the test result appears NaN, it indicates the microphone is not plugged in well or the circuit is not
in the place.
The microphone’s interface goes wrong if the test result appears FAIL.
Suggestion to failure test
Ensure the microphone works well first. The sound waves appear on the screen if the sound
arrives at the microphone The interface does not work if the sound waves do not appear on the
screen. It is recommended to replace the PC carrier board.
C.1.8 Z0207 WIFI Function Test
Top test item
N/A
Test content
Decide whether the wireless network adapter exists, and then search for SSID list.
Analysis to test failure
The wireless network adapter is not recognized if the information “No wireless device found
appears.
Fail to find wireless hotspots if the information “Wireless function test fails” appears.
Suggestion to failure test
Check whether the connection of wireless network adaptor is fixed well and wireless network
adaptor is not forbidden if the information “No wireless device found” appears.
Check whether there is WIFI hotspot if the information “WIFI Function test FAIL appears. If there
is a hotspot, replace the wireless network adaptor.
C.1.9 Z0208 Wired Network Interface Test
Top test item
N/A
Test content
The program informs to plug in the network cable. The program decides the existence of the wired
network adapter, and then decides whether the Ethernet works well.
Analysis to test failure
The Ethernet cannot be recognized if the information “No network adapter found” appears.
The network cable cannot be recognized if the information “No network cable plugged in
appears.
Ethernet does not work well if the test appears FAIL.
Suggestion to failure test
Please check the installation of the network drive if the information “No network adapter found
appears.
Please check the network cable if the information “No network cable plugged in appears.
It is recommended to replace PC carrier board if the test result appears FAIL.
C.1.10 Z0301 DC-DC Voltage Monitor Test
Top test item
PC Module and MF FPGA Interconnection Test.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray Eagus R9 and is the answer not in the manual?

Mindray Eagus R9 Specifications

General IconGeneral
BrandMindray
ModelEagus R9
CategoryDiagnostic Equipment
LanguageEnglish

Related product manuals