System Installation 3-33
NOTE:
1. AE Title should be the same with the SCU AE Title preset in the server
(PACS/RIS/HIS).
2. DICOM communication port should be the same with the one in the server.
3. If the currently entered name has already existed, the system will pop up: “The
server name exists!” Click [OK] to enter another name.
3. Click [Set DICOM Service].
When the system is configured with DICOM basic function module, and installed DICOM Worklist,
MPPS, DICOM Structured Reporting and Query/ Retrieve modules, the corresponding preset
settings can be found in DICOM Service screen.
The DICOM Service Setting is used to set properties of DICOM services as Storage, Print, Worklist,
MPPS, Storage Commitment and Query/ Retrieve. The detailed information please refer to DICOM
chapter in the operator’s manual [Basic Volume].