5-10  Checking Performance and Functions   
5.4.2.3  Cine Review 
Procedure  Checking criteria 
  Press <Freeze> 
system enters manual-review. 
(Precondition: 
Set “Enter in frozen status” 
to “Cine”. 
 
Open cine files in thumbnail, iStation or 
Review. The system enters automatic cine 
review status. 
  Enter Cine Review status 
  The system enters auto review status. 
Roll the trackball  Manual cine review 
Click soft menu-[Auto Review] 
Start auto review. Adjust soft menu knob to 
change the speed of cine review. 
Review speed gets faster with the value 
becoming bigger. 
Review speed gets slower with the value 
becoming smaller. 
If the review speed is 0, it exits auto review mode 
Move the cursor to the start point to review (or 
image frame), and then click soft menu of 
image [Set as Start Point]. 
Set start point of auto review. 
Move the cursor to the end point to review (or 
image frame), and then click soft menu of 
image [Set as End Point]. 
Set end point of auto review. 
Click soft menu-[Auto Review] again. 
The cine review plays between the start point 
and the end point. 
 
Press <Freeze> to defreeze the image 
again. 
 
Return to scan status with the image 
defreezing and exit cine review. 
 
5.4.2.4  Probe Switch and Recognition 
Procedure  Checking criteria 
Press <Freeze> button –  plug the probe –  press 
<Freeze> button – press <Probe> button. 
Ultrasound device can be recognized 
while connecting the probe. 
Press <Freeze> button –  unplug the probe – 
new probe. 
Recognize the probe type instantly. 
 
tart point of 
 
nd point of auto 
 
eview progress bar
 
otal