Part I Service Scenarios
Repair 85
Battery Health
Status
Determining Condition Status Description
WEAK The batteries are free from damage
and recoverable faults.
The full charge capacity of batteries
(FCC) is less than 40% of the rated
capacity.
The number of cyclical recharging
times of batteries is less than 300.
Fully charge batteries and then
discharge them to partially recover the
battery capacity.
REPLACED The batteries are free from damage
and recoverable faults.
The full charge capacity of batteries
(FCC) is less than 40% of the rated
capacity.
The number of cyclical recharging
times of batteries is greater than 300.
The battery capacity deteriorates
gradually after batteries are being used
for a long time. Continuous use will
compromise the power performance
and standby duration of batteries.
Therefore, remind users to replace
batteries in a timely manner.
INVALID Unrecoverable damage occur on
batteries.
If a battery encounters permanent
damage, remind the user to replace the
battery immediately.
PROTECTED Batteries encounter recoverable faults. If batteries encounter recoverable
faults, the batteries will resume
normal after a period of time. If the
problem persists, contact Mindray
Customer Service Department.
3.1.4.2 Abnormal System Voltage
Alarm Message Log Record Recommended Action
The battery life is about to run
out. Replace the battery as
soon as possible.
*** System Monitor: Power
supply alert! [XXX], Current
voltage: [VVV] V, Limit
voltage:[LLL]~[HHH]V
Replace the coin battery.
None "Date" "Time" "Errocode":
System Monitor:Power supply
alert! [XXX], Current voltage
error.
If the system detects that a
voltage is abnormally high or
low, replace the main board.