Part I Service Scenarios
Repair 86
3.1.5 Temperature Error
Alarm Message Log Record
Recommended
Action
"Errocode" FPGA
temperature
warning! Shut
down in xx
seconds.
"Errocode" FPGA
Temperature Alert!
Shut down after xx
Seconds
"Date" "Time" "Errocode": System
Monitor:Temperature alert! [XXX], Current
temperature: [VVV] C, Limit
temperature:[LLL]~[HHH]C.Current fan speed
[SSS] rpm, [SSS] rpm, [SSS] rpm, [SSS] rpm, [SSS]
rpm.
[XXX] indicates the monitoring point name, [VVV]
indicates the current value (monitored value), [LLL]–
[HHH] indicate the upper and lower limits (upper and lower
thresholds of the monitoring point), and [SSS] indicates the
fan speed.
Replace the fan.
"Errocode" CPU
Temperature Alert!
Shut down after xx
Seconds
"Date" "Time" "Errocode": System
Monitor:Temperature alert! [XXX], Current
temperature: [VVV] C, Limit
temperature:[LLL]~[HHH]C.Current fan speed
[SSS] rpm, [SSS] rpm, [SSS] rpm, [SSS] rpm,[SSS]
rpm.
[XXX] indicates the monitoring point name, [VVV]
indicates the current value (monitored value), [LLL]–
[HHH] indicate the upper and lower limits (upper and lower
thresholds of the monitoring point), and [SSS] indicates the
fan speed.
Replace the fan.
3.1.6 Fan Error
Alarm Message Log Record
Recommended
Action
"Errocode" System
Fan requires
maintenance,
please contact the
service!
"Date" "Time" "Errocode": System Monitor: Fan alert!
[XXX], Current speed: [VVV] rpm, Lower Limit speed:
[LLL] rpm.
[XXX] indicates the fan number, [VVV] indicates the
current value, and [LLL] indicates the threshold.
Replace the fan
3.1.7 Key Error
Alarm Message Log Record
Recommended
Action
"Errocode" [XXX] button; it may be blocked.
"Errocode" [YYY] key may be stuck, please
confirm.
[YYY] indicates the key silkscreen.
"Date" "Time"
"Errocode": System
Monitor:Key stuck alert!
[YYY] key is stuck!