1-4 Safety Precautions
10.
Do not allow the patient to contact the live parts of the
ultrasound system or other devices, e.g. signal I/O ports. Electric
shock may occur.
11.
Do not use an aftermarket probe other than those specified by
Mindray. The probes may damage the system causing a
profound failure, e.g. a fire in the worst case.
12.
Do not subject the transducers to knocks or drops. Use of a
defective transducer may cause an electric shock.
13.
Do not open the covers and front panel of the system. Short
circuit or electric shock may result when the system hardware is
exposed and powered on.
14.
Do not use this system when any digital device such as a high-
frequency electrotome, high-frequency therapeutic device or
defibrillator is applied already. Otherwise, there is a risk of
electric shock to the patient.
15.
Only use the ECG leads provided with the physiology module;
otherwise, electric shock may be resulted.
16.
When moving the system, you should hold the handle;
otherwise, damage may be resulted by abnormal force. Do not
push the system from the left/right side, otherwise, it may be
toppled over.
17.
The auxiliary power output outlet in the system is used to supply
power for the recommended peripheral devices. Do not connect
other devices to the outlet, otherwise, the rated output power
may be exceeded and failure may be resulted.
18.
Accessory equipment (analog or digital) connected to the
ultrasound system must comply with the relevant IEC standards
(e.g., IEC 60950 information technology equipment safety
standard and IEC 60601-1 medical equipment standard).
Furthermore, all configurations must comply with the standard
IEC60601-1. It is the responsibility of the person, who connects
additional equipment to the signal input or output ports and
configures a medical system, to verify that the system complies
with the requirements of IEC60601-1. If you have any questions
regarding these requirements, consult your vendor.
19.
Prolonged and repeated use of keyboards may result in hand or
arm nerve disorders for some individuals. Observe the local
safety or health regulations concerning the use of keyboards.
20.
It is not allowed for the operator to have contact with other
patients and the electronic parts (such as the input/output
terminal of the signal) of other devices that are connected to the
system. Otherwise, it may produce the electrical shock to the
patient.
21. When using a probe, pay attention to the status of the ultrasound
image. Do not use the probe to perform image acquisition when the
image is frozen.