Image Optimization 5-99
2. Tap [Elasto]→[STQ] on the touch screen. Or press the user-defined <STQ> key (set the user-defined
key via [Preset]→[System]→[iConsole&Footswitch]/[Key Board]) to enter the STQ mode.
3. Adjust the ROI based on the lesion size and the position.
4. Press <Update> to generate the acquisition. Place the probe still with stable force (not pressing,
sweeping or moving the probe) to acquire the image. Adjust the B image parameters to obtain a
premium image.
The stress curve shows at the bottom of the screen after entering the real-time acquisition.
The following indices display besides end of the curve in real-time:
Elasto modulus inside the ROI of the current frame;
The mean value, maximum value, minimum value, SD value of shear velocity;
Depth value of the active elasto modulus.
5. Freeze the image, and replay the image if necessary.
6. Save the single-frame and multi-frame image.
7. Reacquire the image if necessary.