16-4  EMC Guidance and Manufacturer’s Declaration 
TABLE 3 
GUIDANCE AND MINDRAY DECLARATION—ELECTROMAGNETIC IMMUNITY 
The system is intended for use in the electromagnetic environment specified below. The 
customer or the user of system should assure that it is used in such an environment. 
IMMUNITY 
IEC 60601 TEST 
COMPLIANCE 
ELECTROMAGNETIC 
Conduced RF 
IEC 61000-4-6   
3 Vrms   
0,15 MHz – 80 MHz 
6 Vrms in ISM and 
amateur 
radio bands between 
0,15 MHz and 80 
MHz   
3 Vrms   
0,15 MHz – 80 
MHz 
6 Vrms in ISM 
and amateur 
radio bands 
between 
0,15 MHz and 80 
MHz   
Portable and mobile RF 
communications equipment should 
be used no closer to any part of 
system, including cables, than the 
recommended separation distance 
calculated from the equation 
applicable to the frequency of the 
transmitter. Recommended 
separation distance 
d = 1.2 x
 
d = 1.2 x
  80 MHz to 800 MHz 
d = 2.3 x
  800 MHz to 2.7GHz 
Where, P is the maximum output 
power rating of the transmitter in 
watts (W) according to the 
transmitter manufacturer and d is 
the recommended separation 
distance in meters (m). 
Field strengths from fixed RF 
transmitters, as determined by an 
electromagnetic site survey, should 
be less than the compliance level in 
each frequency range.   
Interference may occur in the vicinity 
of equipment marked with the 
following symbol:   
 
Radiated RF 
IEC 61000-4-3 
10 V/m 
80MHz - 2.7GHz 
10 V/m 
80MHz - 2.7GHz 
Proximity fields 
from RF 
wireless 
communications 
equipment 
IEC 61000-4-3 
27 V/m 
380–390 MHz 
27 V/m 
28 V/m 
430–470 MHz, 800–
960 MHz, 1700–1990 
MHz, 2400–2570 
MHz 
28 V/m 
9 V/m 
704–787 MHz, 5100–
5800 MHz 
9 V/m 
Note 1  At 80 MHz and 800 MHz, the higher frequency range applies. 
Note 2  These guidelines may not apply in all situations. Electromagnetic propagation is 
affected by absorption and reflection from structures, objects and people. 
 Field strengths from fixed transmitters, such as base stations for radio (cellular 
/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast 
and TV broadcast cannot be predicted theoretically with accuracy. 
  To assess the electromagnetic environment due to fixed RF transmitters, an 
electromagnetic site survey should be considered. If the measured field strength in the 
location in which system is used exceeds the applicable RF compliance level above, 
system should be observed to verify normal operation. If abnormal performance is 
observed, additional measures may be necessary, such as reorienting or relocating the 
system. 
  b Over the frequency range 150kHz to 80MHz, field strengths should be less than 