Probes and Biopsy 13-37
(3) Select a proper guiding block and thread the knob of fixing the guiding block through the hole
of installing guiding block, move the guiding block to the desired position, then turn tightly the
knob of fixing the guiding block to fix the guiding block on the support of needle-guided bracket.
(4) Insert a biopsy needle with the same specification as that of the guiding block into the hole of
the guiding block and turn tightly the knob of fixing the needle.
NGB-053/NGB-054
(1) Put on the sterile transducer sheath.
(2) Hold the transducer in one hand, select the correct needle-guided bracket and hold it with the other
hand. Match the inside of the bracket with the outside of the transducer respectively. Mount the
bracket onto the transducer. Rotate the locking nuts of the guided bracket at the right and left sides
to fix the bracket and the transducer. Rotate the needle-shaped adjusting nut to the ultimate position
as shown in the figure.
(3) Hold the transducer. Press the biopsy needle pressure position to separate it from the V-shaped
guiding block. Put the needle into the V-shaped guiding block of the needle guided-bracket.