13-50 Probes and Biopsy
13.3.1 Basic Procedures for Needle Navigation Guiding
Needle navigation guiding biopsy procedure under non-Fusion mode:
1. Connect all accessories of magnetic navigator. Refer to chapter "5.17.2 Magnetic Navigator" and
13.3.2 Installation” for details.
2. Activate needle navigation guiding feature and select needle type and length. Perform accuracy
verification. For details, refer to chapter 13.3.3 Accuracy Verification”.
3. Perform ultrasound image scanning and confirm position/size/structure of tumor/lesion. Mark the
lesion. For details, refer to chapter 13.3.4 Mark”.
Tap [SSC] to perform acoustic speed correction. For details, refer to chapter 5.2.3 B Mode Image
Optimization”.
4. Place the needle next to the probe and adjust the biopsy guideline on the image for recommended
needle position and angle.
The guideline turns green if it is able to reach the lesion mark.
5. Perform biopsy by the guideline.
The distance indicating box shrinks by the time needle tip goes towards the lesion mark. When the
tip reaches the lesion mark, the box turns green and merges with the tip.
For detailed procedure description, refer to chapter13.3.5 Screen Display in the Guiding.
6. Save the single-frame and multi-frame image if necessary.
Needle navigation guiding biopsy procedure under Fusion mode:
1. Connect all accessories of magnetic navigator. Refer to chapter "5.17.2 Magnetic Navigator" and
13.3.2 Installation” for details.
2. Activate needle navigation guiding feature and select needle type and length. Perform accuracy
verification. For details, refer to chapter 13.3.3 Accuracy Verification”.
3. Enter Fusion imaging and load CT/MR/PET/freehand data. Refer to step 1-5 in chapter5.17.4
Basic Proceduresfor details.
4. Mark the tumor/lesion on CT/MR/PET/freehand image. For details, refer to chapter 5.17.6 Marks”.
Tap [SSC] to perform acoustic speed correction. For details, refer to chapter 5.2.3 B Mode Image
Optimization”.
5. Register CT/MR/PET/freehand data. For details, refer to chapter 5.17.4 Basic Procedures”.
6. Place the needle next to the probe and adjust the biopsy guideline on the image for recommended
needle position and angle.
The guideline turns green if it is able to reach the lesion mark.
7. Perform biopsy by the guideline.
The distance indicating box shrinks by the time needle tip goes towards the lesion mark. When the
tip reaches the lesion mark, the box turns green and merges with the tip.
For detailed conversion procedure description, refer to chapter 13.3.5 Screen Display in the
Guiding”.
8. Save the single-frame and multi-frame image if necessary.