EasyManuals Logo

Mindray VS-800 Service Manual

Mindray VS-800
74 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #34 background imageLoading...
Page #34 background image
Product Specifications
3-2
3.2.2.3 Power Requirements
Voltage: AC 100 - 240 V, 50/60 Hz (± 3 Hz)
Power: 70VA
Switch : Standby
3.2.3 Specifications of Display
LCD display
Dot matrix: 320×160
Dot pitch: 0.24×0.24
Dot size: 0.225×0.225
LCD type: FSTN LCD
Polarity: Positive
Transmission mode: Transflective
Viewing angle: 12:00
Type of backlight: LED (4 white LEDs)
Displayed contents:
Startup screen
Trend data (including scroll bar)
SpO
2
waveform (can be switched)
Menu (pop up)
Time
Prompting messages
6 groups of LED digit displays:
Used to display the systolic pressure, diastolic pressure, mean pressure/cuff pressure,
SpO
2
, PR and TEMP.
8 groups of monochrome LED indicators
Used to display or indicate the pulse strength, power, ADULT-PEDIATRIC-NEO (patient
type), mmHg-Kpa (unit of pressure), silence function, NIBP measurement status,
ºC -
ºF
(temperature unit) and indication lamps which indicate temperature type
.
3.2.4 Signal Interface
Nurse call output The nurse call function is designed for the
nurse call system in hospital
Type of nurse call signal Relay signal, always ON/OFF (optional)
Maximum voltage 12V DC (no greater than 13V DC)