Product Specifications
3-3
Maximum load current ≤2A
Connection resistance <1Ω
Isolation voltage >1500V AC
Network interface For connection to the Central Monitoring
Network, software upgrade and data export
3.2.5 Battery
No. of batteries: 1
Battery type: Sealed lead-acid battery or lithium ion battery
Time to shutdown at low battery power: 5-15min after the first low-battery alarm (a new
fully-charged battery should be used)
3.2.5.1 2.3 Ah 12V Sealed Lead-Acid Battery
Minimum of 260min
The monitor runs on power supplied by the new fully-charged battery in the following
conditions:
Ambient temperature: 25℃
Monitor configuration: SpO
2
(Continuous measurement) and NIBP (one auto NIBP
measurement per 15min)
Charge time: Maximum of 8h. Operating status.
3.2.5.2 4.4Ah 11.1V Lithium Ion Battery
Minimum of 620min
The monitor runs on power supplied by the new fully-charged battery in the following
conditions:
Ambient temperature: 25℃
Monitor configuration:
SpO
2
(Continuous measurement) and NIBP (one auto NIBP
measurement per 15min)
Charge time: Maximum of 8h. Operating status.
3.2.6 Recorder
Method Thermal dot array
Horizontal resolution 160dots/cm (25 mm/s)
Vertical resolution 80dots/cm
No. of waveform channels 1
Paper width 50±0.1 mm