MBD-2 Operating Manual 26
15-00167, Rev 1
19 Too early REG_UP
Problem: The REG_UP I/O line signal was detected too early
due to a failure in the voltage block regulator.
Resolution: Restart device by removing and re-inserting
battery. If error message repeats, remove device from service.
1B
Factory parameters
applied
Problem: The measurement result memory does not contain a
valid measurement result record due to empty or corrupt flash
memory
Resolution: Restart device by removing and re-inserting
battery. If error message repeats, remove device from service.