18(21)
6. RDS-30 specifications
Order number RDS-30 µSv/h: 1233-244
Order number RDS-30 mrem/h: 1233-250
6.1 Physical characteristics
radiation detected:
gamma and x-rays, 48 keV...1.3 MeV
detector:
energy-compensated GM tube. Energy response according to ambient
dose equivalent H*(10)
measurement range:
dose rate 0.01 – 100 000 µSv/h or 0,001 – 10 000 mrem/h
dose 0.01 – 1 000 000 µSv or 0,001 – 100 000 mrem
alarm levels:
adjustable alarm levels for dose rate and accumulated dose
calibration accuracy:
± 5 % of the reading in
137
Cs exposure, at 3 mSv/h (300 mrem), +20°C (68°F)
dose rate linearity:
±10 % ± 1 digit in the range of 0.0001 - 100 mSv/h
response to radiation energy:
±30 % over the range of 48 keV - 1.3 MeV
response to angle of radiation incidence:
±25 % within ± 45° from the calibration direction with 48 keV energy
6.2 Electrical characteristics
power supply:
2 x alkaline cell IEC LR6/AA
battery life:
at least 2000 hours at normal background with alkaline cells
battery alarm:
two-step alarm for low battery voltage
resistant against RF interferences
6.3 Mechanical characteristics
case:
rugged plastic case
dimensions:
78 x 126 x 32 mm
weight:
170 g without batteries
220 g with batteries
6.4 Environmental characteristics
temperature range:
-25...+55°C operational (-13°F to 131°F)
-40...+70°C storage (-40°F to 158°F)
6.5 Connection
IrDA link to PC