2.6.7 Error (dEF/Err)
The display alternates between the texts "dEF"
(defective) and "Err+number". The continuous
audible alarm is on.
Error alarm:
_____________
(continuous beep). The
dose rate measurement is prevented.
If the "Err+number"-text alone is alternating
with the dose rate display, the dose rate measurement is continued.
Only a visual alarm is given, no audible alarm.
Error codes and types:
Error1 = Calibration
Error2 = Non-volatile memory
Error3 = Detector
Error4 = High voltage
Error5 = Internal power supply
Error6 = Cumulative dose overflow (>10 Sv)
Should an error occur, write down the Error number (Err + number) and return the
meter to the manufacturer (contact info: see last page of this manual) for servicing.