Operation Manual _____________Programming the Microprocessor of XL2020
_________________________________________________________________________________________________
8-2
CONTINUOUS MODE
Ultrasonic energy is delivered continuously without any timing or pulsing cycles. The
ultrasonic output will shut down when the total processing time reaches 9:59:59
(hr:min:sec).
Programming Steps and Descriptions Action LCD Display
Turn ON generator.
Press ā€œIā€ on Power
Switch
SONICATOR XL2020
MISONIX INCORPORATED
(Scrolls on screen, then switches to)
For Tuning Procedure
Refer to Manual
(alternates with)
If SONICATOR is Tuned,
Press CLEAR
Clear introduction message.
Press CLEAR Key
Press PROG to Program
Timer and PULSAR or
(alternates with)
Press START for
Continuous Operation
Start sonication in continuous mode.
(Turn Amplitude Control Knob to desired
level)
Press START Key
STOP
Continuous 0:00:00
(Elapsed Time begins to count seconds)
To interrupt processing. Press PAUSE Key Continuous X:XX:XX
Press PAUSE to Continue
To continue processing. Press PAUSE Key Continuous X:XX:XX
(Timer continues to count seconds)
Stop sonication when finished. Press START Key
STOP
Press PROG to Program
Timer and PULSAR or
(alternates with)
Press START for
Continuous Operation