3 Setting Parameters
11
No. 99MBC109B
3
Setting Parameters
The settings of the Linear Gage that you will use, the display of the Counter,
and external output are specied by setting parameters. Set parameters before
you begin measuring.
2-axis models have 4 internal counters that are referred to as "CEL", and
depending on the Display mode setting, 6 types of counter values can be
specied for each CEL separately. For details, see
"3.4 Details of Display
Mode (2-axis Models Only)" (page 17).
3.1
Procedure for Setting Parameters
Parameters are set in Parameter mode. As an example, the procedure for using
the Linear Gage with a resolution of 5 µm for EH-102P is explained.
1
Turn on the power switch.
» The Counter enters the stand-by
state.
2
Press and hold [P.SET], and then press [A_ZERO]/[ZERO].
» The Counter enters Parameter mode.
Parameter number Set value
3
Press [P.SET] repeatedly to advance the parameter number to 12.
» The current value of parameter
number 12 of INPUT A will be
displayed.
(Parameter number 12 sets the
resolution.)
Parameter number
INPUT
number
Set value
4
Repeatedly press [A_ZERO]/[ZERO] to set the set value to 1 (resolution:
5 µm).
» The value of INPUT A will be set to 1
(the Linear Gage resolution of INPUT
A will be set to 5 µm).
Parameter number
INPUT
number
Set value