3 Setting Parameters
12
No. 99MBC109B
5
Press [P.SET].
» The current value of parameter
number 12 of INPUT B will be
displayed.
Tips
For 1-axis models, the parameter number will advance. Proceed to step
8
.
6
Repeatedly press [A_ZERO]/[ZERO] to set the set value to 1 (resolution:
5 µm).
» The value of INPUT B will be set to 1
(the Linear Gage resolution of INPUT
B will be set to 5 µm).
Parameter number
INPUT
number
Set value
7
Press [P.SET].
»
The set value will be applied.
8
Press and hold [P.SET], and then press [A_ZERO]/[ZERO].
» The Counter will return to the stand-
by state.
3.2
Basic Parameters
This section explains the basic parameters related to measuring. Be sure to set
these settings before measuring.
Tips
Correct measurement results may not be obtained if the settings are incorrect.
The circles in the Counter model columns in the following table indicate whether a
parameter number that is displayed on the Counter is valid or invalid (●: valid, ○: invalid).
No. Setting item
Description/Allowable values
(the values in bold indicate the default value)
Counter model
P Z S D
00
Parameter mode Used for setting/saving/loading parameters or setting
the optional constant value.
0: Parameter setting
1: CEL-specic parameter setting
2: Constant value setting
3: Save parameters
4: Load parameters