6 Troubleshooting
51
No. 99MBC109B
Display/
Total tolerance
indicator
RS-232C
output
NOM
signal
Tolerance/
BCD
Cause
Solution/
Error cancellation method
Error 55
Red indicator
on
H L1 = L
L5 = L
FFFF55
RS LINK
malfunction,
HOLD trigger
malfunction
Check the unit's connection
status, I/O connector connection
circuit, and power, etc.
Turn on the power switch again.
Error 70
Red indicator
on
Error_70 H L1 = L
L5 = L
FFFF70
Wrong resolution
for calculation
axis
Revise the measurement
conditions.
Automatic cancellation
Error 80
Red indicator
on
Error_80 H L1 = L
L5 = L
FFFF80
Peak detection
error
(Excess speed)
Revise the measurement
conditions.
y
[SEL]/[CE]
y
CS00CRLF (RS)
y
HOLD input (I/O)
Error 90
Error_90 L Counter
value status
Tolerance value
setting error
Input the tolerance value again.
Counter
value status
[SEL]/[CE]
Error 95
Normal
output
L Counter
value status
Key protect Cancel the key protection
parameters.
Counter
value status
Automatic cancellation
Tips
In the "Solution/Error cancellation method" column, "RS" denotes an RS232C command
and "I/O" denotes an external HOLD signal.
The error output format of RS-232C is "CH**,Error$$CRLF".
"Error 40" (Linear Gage malfunction) is displayed even when no Linear Gage is
connected.
"Error 90" and "Error 95" are displayed when a tolerance value setting error occurs due to
a key operation.
"Error_90" in the RS-232C output means that a tolerance value setting error has occurred
due to an RS command.
If an error occurs while you are setting parameters, the Preset value, or the tolerance
value, the error will be displayed after you return to the counter state. However, the error
signal will be output immediately to any external output.