20
2 Basic Measuring Methods
No. 99MAF600B
5
Measure the bottom surface of the ball diameter calibration block.
1 Move the ball diameter calibration block (or main unit), and move the stylus to the following position near the
ball diameter calibration block.
2 Turn the probe unit up/down wheel, and slowly bring the stylus into contact with the bottom surface of the
ball diameter calibration block until there is a "beep" sound.
2
1
2
» The probe diameter calibra-
tion is complete.
Tips
The dimensions of the ball diameter and the calibrated value of the probe diameter may not match exactly
due to bending of the stylus.
The moment the beep sounds, the displayed value may appear to jump. The calculation of the ball's diam-
eter will then be displayed.
6
Press [ENTER].
» The ball diameter setup is complete.