4.1 If the Following Problem Occurs
Based on the problem, determine the cause and the solution.
Problem Cause Solution
The power does not
turn on when [Power
button] is pressed.
The batteries are not inserted
correctly.
Reinsert the batteries.
The batteries have died. Replace the batteries with new ones.
The AC adapter (optional ac-
cessory) is not properly con-
nected.
Reconnect the AC adapter (optional acces-
sory).
The batteries are
depleted quickly.
The LED is set for continuous
use.
y Change the setting of the LED lighting.
Tips
For details, see "3.7 Setting the LED Light-
ing Time" on page 50.
y Remove the batteries if the product will not
be used for an extended period of time.
Theair-oatfunction
does not work (the
battery depletion
icon is only lit when
theair-oatswitchis
pressed).
The batteries are too low for
theair-oatfunctiontooperate.
y Replace all four batteries with new ones.
y Connect an AC adapter (optional acces-
sory).
Tips
Theair-oatfunctionusesanexcessive
amount of electrical power. If you use this
function often, using an AC adapter (op-
tional accessory) is recommended.
Whenusingtheair-oatfunction,the
batterylife(untiltheair-oatfunctionno
longer operates) is around 1.5 hours.
Measurements are
not made even when
the tip of the stylus
comes in contact with
a workpiece (there is
no beep sound).
The clamp knob is locked. Loosen the clamp knob.
The displayed value
is being held, and
the workpiece cannot
be measured ([H]
remains displayed on
the LCD).
The displayed value is held. Press the [HOLD DATA] key to release the
hold on the value.
The clamp knob is locked. Followtheinstructionsin"■Whenthecount-
er value is held, and the measurement can-
not be made" on page 58 to loosen the
clamp knob.