EVA 3 Plus: Device User Manual
6. Care and Maintenance
6.1 Storage
We recommend leaving the colposcope in its provided case for protection and
to minimize dust contamination when not in use. Always follow the cleaning
process outlined below prior to returning the unit to its case to avoid
contamination.
6.2. Lens Care
The optical unit lens should be kept free from loose debris or dust particles
that may block the field of view. In the event of dirt or smudges on the lens
surface, clean the lens with a soft cloth or lens wipe. We recommend using
the cloth provided in your care kit. Never use towels.
6.3. Body/Base
The optical unit needs to be cleaned between each and every use due to the
colposcope accumulating dirt over time, and the possibility of
cross-contamination.
6.4 Cleaning Instructions
Cleaning the base of the optical unit should only be done with CaviWipes
disinfectant towelettes, or their equivalent.
Use 3-5 towelettes, at minimum, to wipe the optical unit
body from the handle, both down the internal side facing the
user and the external side facing the patient, making your
way to the top of the unit.
This should be done in a scrubbing motion, covering both
sides of the plastic body.
Be sure to cover all areas and crevices so that the entire
surface body of the unit is wiped down.