EVA 3 Plus: Device User Manual
9. Electromagnetic compatibility - manufacturer declaration
● The EVA System requires special precautions in regard to
electromagnetic compatibility.
● It must be installed and prepared for use as described in the
Operation section of this User Manual.
● The separation distances recommended in this paragraph must
therefore be complied with.
● The EVA System must not be used near or on top of another
electronic device. If this cannot be avoided, it is necessary – before
clinical use – to check the equipment for correct operation under the
conditions of use.
● The use of accessories other than those specified or sold by
MobileODT as replacement parts may have the consequence of
increasing the emissions or decreasing the immunity of the unit.
9.1. Electromagnetic emissions
● The EVA System is intended for use in the electromagnetic
environment specified in the following tables 1, 2, 4 and 6 below.
● The user and/or installer of the unit must ensure that it is used in
such an environment.
The EVA System 3 Plus is intended for use in the electromagnetic environment specified below; The customer
or the user of the EVA System 3 Plus should assure that it is used in such an environment.
The EVA System 3 Plus uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not likely to cause any