9. Electromagnetic compatibility - manufacturer declaration
The EVA System requires special precautions in regard to
electromagnetic compatibility.
It must be installed and prepared for use as described in the
Operation section of this User Manual.
The separation distances recommended in this paragraph must
therefore be complied with.
The EVA System must not be used near or on top of another
electronic device. If this cannot be avoided, it is necessary – before
clinical use – to check the equipment for correct operation under the
conditions of use.
The use of accessories other than those specified or sold by
MobileODT as replacement parts may have the consequence of
increasing the emissions or decreasing the immunity of the unit.
9.1. Electromagnetic emissions
The EVA System is intended for use in the electromagnetic
environment specified in the following tables 1, 2, 4 and 6 below.
The user and/or installer of the unit must ensure that it is used in
such an environment.
Table 1
Guidance and manufacturer's declaration – electromagnetic emissions – EVA System 3 Plus
The EVA System 3 Plus is intended for use in the electromagnetic environment specified below; The customer
or the user of the EVA System 3 Plus should assure that it is used in such an environment.
Emissions test
Compliance
Electromagnetic environment - guidance
RF emissions
Group 1
The EVA System 3 Plus uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not likely to cause any
37