EVA 3 Plus: Device User Manual
CISPR 11
interference in nearby electronic equipment.
RF emissions
CISPR 11
Class B
The EVA System 3 Plus is suitable for use in all establishments, including
domestic establishments and those directly connected to the public
low-voltage power supply network that supplies buildings used for
domestic purposes.
Immunity test
IEC 60601-1-2
Test level
Compliance level
Electromagnetic environment - guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
8 kV contact
15kV air
8 kV contact
15 kV air
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30 %.
Power
frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
30 A/m
30 A/m
Power frequency magnetic fields should be at
levels characteristic of a typical public
low-voltage power supply network that
supplies buildings used for domestic purposes,
commercial or hospital and clinical
environments.
Table 4