Guidance and manufacturer's declaration – electromagnetic immunity – EVA System 3 Plus
The EVA System 3 Plus is intended for use in the electromagnetic environment specified below; the
customer or the user of the EVA System 3 Plus should ensure it is used in such an environment.
Immunity test
IEC 60601-1-2
Test level
Compliance level
Electromagnetic environment -
guidance
Radiated RF
IEC 61000-4-3
10 V/m
80 MHz to 2.5 GHz
10 V/m
Portable and mobile RF
communications equipment
should be used no closer to any
part of the
EVA System 3 Plus, including
cables, then the
recommended separation
distance calculated from
the equation, applicable to the
frequency of the
transmitter.
Recommended separation
distance
d
=6/EP 80 M Hz to 2.5 GHz
where P
is the maximum output
power rating of the
transmitter in watts (W)
according to the transmitter
manufacturer and d
is the
recommended separation
Distance in meters (m) and E is
the immunity test level in V/m.
Field strengths from fixed R F
transmitters, as
determined by an
39