OPERATOR'S MANUAL
17
WARNING
IT IS ADVISABLE TO CONNECT THE DRAINAGE PIPE TO A SIPHON
WITH A MINIMUM HEIGHT OF 50 MM THAT CAN BE INSPECTED FROM
THE LOWER PART.
During installation operations please observe the following precautions:
- since the temperature of drainage water is 95°C, both drainage pipe ends must be fixed
to the hose fitting and duly secured in place with the supplied clamp;
- the end of the drainage pipe must be placed at a height that complies with specifications,
with respect to the support surface of the device;
- the inner diameter of the fixed drainage pipe shall be at least 40 mm;
- prolongations of the supplied drainage pipe shall not be carried out. Possible
prolongations may cause drainage problems.