OPERATOR'S MANUAL
33
DECALCIFICATION CYCLE
A white thin layer could deposit on cover and chamber metal surfaces due to a high setting of
water hardness value.
Machine can be cleaned running the special "Decalcification" cycle.
When carrying out this cycle, use a special limescale remover for water fittings, to be filled in
manually. Use only limescale removers for water fittings based with Sulphamic Acid and/or
Citric Acid and/or Phosphoric Acid.
WARNING
AVOID USING TOILET PRODUCTS CONTAINING HYDROCHLORIC ACID
OR NITRIC ACID OR SULPHURIC ACID AS THEY COULD DAMAGE THE
DEVICE.
Start the cycle without pouring product inside chamber. Wait for the next message and pour
product in.
When requested, pour 50 ml of product inside chamber and close cover.
NOTE
BEFORE PRESSING "YES", MAKE SURE THAT PRODUCT HAS BEEN
POURED INSIDE CHAMBER AND THAT COVER HAS BEEN CLOSED.