46
OPERATOR'S MANUAL
For cycle correct setting, set thermal disinfection temperature (80°C or 90°C), A0 value,
washing duration and drying duration.
A0 parameter can be set between 600 and 6000 with increments of 600, with thermal
disinfection at 90°C.
If thermal disinfection at 80°C is selected, A0 parameter can be set between 600 and 1200, in
steps of 300.
After all these selections the estimated total time appears.
Use key to save settings and go back to the previous screen.
Press START key to start the cycle defined by the user.
NOTE
THE A0 VALUE SET CORRESPONDS TO THE MINIMUM ENSURED FOR
EACH DISINFECTION PROGRAMME.
ONCE CYCLE IS COMPLETED, THE REACHED AND DISPLAYED A0
VALUE WILL BE HIGHER THAN THE ONE SHOWN UPON CYCLE START.
NOTE
SETTINGS USED ARE STORED FOR A POSSIBLE FOLLOWING USE OF
CUSTOM CYCLE.
WARNING
AT THE END OF THE CYCLE IT IS ESSENTIAL TO VISUALLY INSPECT
THE OBJECTS TO CHECK THEIR PROPER WASHING.