Failure to control board.
If the problem persists, contact
Technical Service Department.
Failure to control board.
If the problem persists, contact
Technical Service Department.
Failure to control board.
If the problem persists, contact
Technical Service Department.
Failure to control board.
If the problem persists, contact
Technical Service Department.
Replace the USB stick and repeat
the download operation.
Non-compatible or failed USB stick.
Ensure that the USB stick is
operative, by inserting it in a PC.
Check that formatting is correct (see
'Technical data' section). If the
problem persists, try with a different
USB stick.
If the problem still persists, contact
Technical Service Department.
USB stick with insufficient space or
write-protected. Too many files in the
main directory.
Remove writing protection or free up
memory (at least in part) on USB
stick.
If the problem persists, contact
Technical Service Department.
Failure to control board.
If the problem persists
contact Technical Service
Department.
Cycle report to be downloaded
recommended limit has been
reached.
Download cycle report through the
"New" option.
Cycle report to be downloaded limit
(500) in case of SD Card malfunction
has been reached. If reports are not
downloaded, data will be overwritten
and permanently lost.
Download cycle report through the
"New" option.
If the problem persists, contact
Technical Service Department.
Malfunction in software running.
Check, using a watchdog, that one of
main processes is not in lock
condition.
If the problem persists, contact
Technical Service Department.
Malfunction in software running.
Check, using a hardware watchdog,
that one peripheral is not in lock
condition.
If the problem persists, contact
Technical Service Department.
Malfunction in software running.
Check, using a watchdog, that one of
main processes is not in lock
condition (e.g. infinite loop).
If the problem persists, contact
Technical Service Department.
Malfunction in software running
If the problem persists, contact
Technical Service Department.
Malfunction in software running.
Restart the thermo-disinfector, if the
problem persists contact Technical
Service Department.